Maintenance
Fault finding
Try another battery set (safely)
Remove & replace battery clip
(safely)
Replace batteries (safely)
Back light drains battery
life
Excessive use of backlight
Avoid or reduce use of backlight
Software locked out by
hardware
Remove & replace batteries (safely)
Contrast control setting wrong
Turn off & on; press & hold contrast
Bar graph/number do not
change
Press zero or change Peak Hold
state
Range not set in Auto-range
Press & hold Range until Auto-range
He range will not show –5
Unit with nozzle ON and/or
range
Reset via Nozzle and Range keys
Gas Group not set correctly
Press & hold Gases until shown OK
Nozzle/Peak is not set
correctly
Press & hold Nozzle until both
correct
Unit has exceeded 1 year on
cal
Return to Ion Science for
recalibration
Unit will not zero correctly
Probe blocked by solid or
liquid
Remove probe; clean; check now OK
Motor failed & probe not
blocked
Return to supplier for repair
Instrument readings noisy
Instrument readings noisy
Test environment
contaminated
Move test into clean area
5000is will not switch on
Return to supplier for repair